EM5040 Series LISN
Line Impedance Stabilization Networks for EMC Testing
The EM5040 Series is a family of V-type line impedance stabilization networks designed for conducted disturbance voltage measurement in EMC testing. The series supports R&D debugging, compliance validation, and industrial EMC workflows with models optimized for standard CE testing, receiver protection, CM/DM analysis, and high-current single-line applications.
Stable Impedance for Accurate Conducted Emission Measurement
The EM5040 Series is a family of V-type line impedance stabilization networks designed for conducted emission testing and EMC debugging. These LISNs provide a stable impedance environment for the equipment under test while coupling disturbance signals to external receivers or analyzers for accurate measurement.
The series includes multiple configurations for different workflows. EM5040E provides the standard single-phase LISN function, EM5040A adds a built-in 10 dB limiter and optional 150 kHz high-pass filter, EM5040B adds common-mode and differential-mode separation for advanced debugging, and EM5040C targets high-current, higher-frequency single-line testing for industrial and automotive-related applications.
Model Selection Guide
| Model | Positioning | Key Function | Best Fit |
|---|---|---|---|
| EM5040E | Standard single-phase LISN | Basic V-type LISN for conducted disturbance voltage measurement | General CE testing and baseline EMC measurement |
| EM5040A | Protected test setup LISN | Built-in 10 dB limiter and optional 150 kHz high-pass filter | Receiver protection and compliance-oriented workflows |
| EM5040B | Advanced EMI debugging LISN | CM/DM separator for common-mode and differential-mode analysis | EMI filter debugging and interference root-cause analysis |
| EM5040C | High-current / high-frequency LISN | 100 kHz to 200 MHz range and 100 A rating | Industrial, automotive, and high-current EMC test scenarios |
EM5040E — Standard LISN
| Parameter | Value |
|---|---|
| Frequency Range | 9 kHz to 30 MHz |
| Circuit Type | (50uH + 5Ω) ‖ 50Ω Type V |
| Measurement Circuit | Single phase two wire L/N |
| AC Voltage | 0 to 264 VAC, 50/60 Hz ±5% |
| DC Voltage | 0 to 375 VDC |
| Current Rating | 16 A |
| Output Connector | 50Ω BNC female |
| Artificial Hand | Yes |
| Weight | 5 kg |
EM5040A — Protected LISN
| Parameter | Value |
|---|---|
| Frequency Range | 9 kHz to 30 MHz |
| Circuit Type | (50uH + 5Ω) ‖ 50Ω Type V |
| Measurement Circuit | Single phase two wire L/N |
| AC Voltage | 0 to 264 VAC, 50/60 Hz ±5% |
| DC Voltage | 0 to 375 VDC |
| Current Rating | 16 A |
| Output Connector | 50Ω BNC female |
| Limiter | 10 dB, 130 dBuV threshold |
| High-Pass Filter | Optional 9 kHz / 150 kHz |
| Weight | 5.25 kg |
EM5040B — CM/DM LISN
| Parameter | Value |
|---|---|
| Frequency Range | 9 kHz to 30 MHz |
| Circuit Type | (50uH + 5Ω) ‖ 50Ω Type V |
| Measurement Circuit | Single phase two wire L/N |
| AC Voltage | 0 to 264 VAC, 50/60 Hz ±5% |
| DC Voltage | 0 to 375 VDC |
| Current Rating | 16 A |
| Output Connector | 50Ω BNC female |
| CM/DM Separator | Yes |
| Weight | 5.25 kg |
EM5040C — High-Current LISN
| Parameter | Value |
|---|---|
| Frequency Range | 100 kHz to 200 MHz |
| Circuit Type | (5uH + 1Ω) ‖ 50Ω |
| Measurement Circuit | Single wire |
| AC Voltage | 0 to 250 VAC, 0 to 400 Hz |
| DC Voltage | 0 to 600 V |
| Current Rating | 100 A rated / 500 A short-time |
| Output Connector | 50Ω N female |
| Weight | 1.8 kg |
Typical Applications
- Conducted disturbance voltage measurement — standard CE testing per CISPR 16-1-2
- EMC compliance pre-test and formal test preparation — receiver-protected workflows
- EMI filter debugging in switched-mode power designs — CM/DM separation for root-cause analysis
- Common-mode and differential-mode interference analysis — advanced filter characterization
- Industrial and automotive EMC validation — high-current high-frequency testing
Key Advantages
- Standards-OrientedDesigned to meet CISPR16-1-2:2006 requirements for V-network conducted emission testing
- Stable ReferenceProvides consistent impedance for the DUT while isolating unwanted mains interference
- EMI Debug ReadySelected models support common-mode and differential-mode analysis for filter debugging
- Receiver ProtectionLimiter-equipped model helps protect sensitive receivers and spectrum analyzers
- Series FlexibilityCovers standard single-phase, protected, CM/DM, and high-current high-frequency use cases
Performance Highlights
- ChannelsSingle-phase L/N
- Power16 A / 100 A models
- Key Spec9 kHz – 200 MHz
- TypeV-type LISN
Request a Demo
See the EM5040 Series LISN in action. Our engineers provide 1-on-1 consultations for your specific testing needs.
Get a QuoteLooking for EMC test equipment or a custom LISN solution?
XGY Tek specializes in power-line stabilization and EMC validation setups for conducted emission measurement, EMI debugging, and compliance testing.