Frequency range
DC–40 GHz
23 ± 3 °C; DC to 10 MHz referenced to 10 MHz, not separately tested
Open-standard phase offset
±2.0°
23 ± 3 °C; relative to reference standards
Fixed-load VSWR
≤1.07
23 ± 3 °C
Adapter VSWR
≤1.12
23 ± 3 °C

Overview

MCAL-KIT-CY 40 is a coaxial mechanical calibration kit for vector network analyzers. The package includes male and female 2.92 mm open, short and load standards, together with adapters and USB calibration data.

Electrical specifications apply at 23 ±3 °C. DC to 10 MHz performance is referenced to 10 MHz and is not separately tested; open- and short-standard phase offsets are relative to reference standards.

Additional product details

MCAL-KIT-CY 40 is a coaxial calibration kit for vector network analyzer measurements.

Calibration Kit — Family Illustration

The image is a family illustration for appearance only. It does not identify an exact model or prove connector geometry or component quantities. Refer to the specifications and standard delivery contents.

Use and care

Take electrostatic discharge precautions before use.

Inspect the connector interfaces before use. Remove dust, cotton fibres, metal fragments and other contaminants, which can substantially affect electrical performance.

Check that the mating connector pin depth is within its normal limits before connection. Abnormal pin depth can permanently damage the calibration standards.

Use a torque wrench to tighten the coupling nut whenever possible to ensure proper operation.

When connecting a standard, rotate the coupling nut on the male connector. Do not rotate the calibration standard or adapter; doing so can sharply reduce its service life.

Store the components carefully and protect them from dust, moisture and severe vibration. If a component is dropped, have its performance verified by a qualified calibration laboratory before returning it to service.

Arrange periodic metrological calibration so that the kit remains within specification during use.

Key Capabilities & Applications

Calibrate both connector genders

The supplied 2.92 mm male and female open, short and fixed-load standards support coaxial calibration. Female–female, male–male and male–female adapters are included.

Apply the calibration data

A USB drive contains calibration data. Open and short phase offsets are ±2.0° relative to reference standards, with electrical specifications applying at 23 ±3 °C.

Protect the standards in use

Inspect connector cleanliness and pin depth before connection. Tighten the coupling nut using the supplied torque wrench and arrange calibration verification after a component is dropped.

Common applications

  • Vector network analyzer coaxial calibration

Technical Specifications

ParameterSpecification
Electrical Specifications
Frequency rangeDC–40 GHz23 ± 3 °C; DC to 10 MHz referenced to 10 MHz, not separately tested
Open-standard phase offset±2.0°23 ± 3 °C; relative to reference standards
Short-standard phase offset±2.0°23 ± 3 °C; relative to reference standards
Fixed-load VSWR≤1.0723 ± 3 °C
Adapter VSWR≤1.1223 ± 3 °C
Environmental Specifications
Operating temperature+20 °C to +26 °C
Storage temperature−20 °C to +70 °C
Performance notes

Electrical specifications apply at an operating temperature of 23 ± 3 °C.

Specifications from DC to 10 MHz are referenced to 10 MHz and are not tested separately.

Open- and short-standard phase offsets are relative to reference standards.

Specifications are subject to change without notice.

Models, Configuration & Compatibility

MCAL-KIT-CY 40

Standard delivery contents

  • 2.92 mm male open standard (×1)
  • 2.92 mm female open standard (×1)
  • 2.92 mm male short standard (×1)
  • 2.92 mm female short standard (×1)
  • 2.92 mm female load standard (×1) — Fixed-load standard
  • 2.92 mm male load standard (×1) — Fixed-load standard
  • 2.92 mm female–female adapter (×1)
  • 2.92 mm male–male adapter (×1)
  • 2.92 mm male–female adapter (×1)
  • 3.5 mm torque wrench (×1)
  • USB drive containing calibration data (×1)
Model selection details

MCAL-KIT-CY 40 is the DC–40 GHz coaxial mechanical calibration kit.

Technical Resources

Download Datasheet

PDF technical specifications

Get a Quote

Pre-filled product inquiry

Selection guidance

Format: Mechanical coaxial calibration kit

Coverage: Frequency range DC–40 GHz at 23 ± 3 °C, DC to 10 MHz referenced to 10 MHz and not separately tested; 2.92 mm standards

Best for: Vector network analyzer coaxial calibration.